1. Home
  2. GLUCUBE APP

Warnings and errors in GLUCUBE measurement

 Possible solutions to warnings and errors that may occur in GLUCUBE measurement

When taking a measurement with GLUCUBE device, various problems may occur due to incorrect use of the device.


If after taking the measurement a green screen does not appear showing a glucose value, this is because there has been an error during the measurement.

Warnings and errors are coded and will appear in the upper left part of the screen:

IMG_20241004_095517

The different error categories that may appear are:

Category E1:

If this error code appears, follow these instructions:

  • Press exit
  • Restart the phone
  • Turn on the device and repeat the measurement process

If the error persists, please contact Customer Service

Category E2

E2-1

This message may appear when starting a GLUCUBE measurement.

The possible causes for this error are:

  • The maximum waiting time between turning on the device and inserting the finger has passed.

The solutions for this type of error are:

  • Press cancel and repeat a measurement, this time making sure that no more than 10 seconds pass between turning on the device and inserting the finger.

If the problem persists, please contact Customer Service 

E2-2

This message may appear during STEP 1 during the signal check step.

The possible causes for this error are:

  • The finger was not in the correct position when preparing the measurement and therefore the measurement has not started.

The solutions for this type of error are:

  • Take another measurement, making sure that the finger perfectly covers the window inside the device without applying pressure.
  • Follow the instructions for use of the device and place your finger correctly

If the problem persists, please contact Customer Service

E2-3 ... E2-5

If this error code appears, follow these instructions:

  • Press exit
  • Restart the phone
  • Turn on the device and repeat the measurement process

If the error persists, please contact Customer Service 

Category E3

E3-1 ... E3-9

This message appears after having performed the GLUCUBE measurement.

The possible causes of this error are:

  • Slight movements that have caused instability in the signal
  • Forced breathing that may have caused instability in the signal
  • Very cold hands that do not allow a correct signal to be detected
  • Taking the measurement on an unstable surface
  • Using nail polish or acrylic nails that may make it impossible to detect the signal

The solutions for this type of error are:

  • Keep the hand as stable as possible, avoiding movements
  • Avoid forced positions that prevent the correct flow of blood in the hand
  • Take the measurement on a stable surface
  • Breathe calmly, avoiding forcing breathing
  • Make sure that the finger correctly covers the window and is in a correct position following the instructions for use of the device

If the problem persists, please contact Customer Service

E3-10

This message appears after having taken the GLUCUBE measurement.

The possible causes of this error are:

  • Changes in the signal captured by the device compared to the previous measurement, which has affected the estimation of blood glucose levels.

The solutions for this type of error are:

  • Perform a new measurement to obtain a more accurate result.
  • Follow the instructions for use of the device and make sure that the finger correctly covers the window and is in a correct position.

If the error persists, please contact Customer Service.

E3-100

The possible causes for this error are: 

  • Communication failure with the cloud service

The solutions for this type of error are:

  • Press “Exit” and turn the device back on to perform a measurement.

If the problem persists, please contact Customer Service 

E3-101

The possible causes for this error are:

  • Problems in data transmission and decompression

The solutions for this type of error are:

  • Change the hardware configuration to "adaptive" mode from the "About the device" section

If the problem persists, please contact Customer Service


E3-102

The possible causes for this error are:

  • Incomplete or incorrectly filled profile data

The solutions for this type of error are:

  • Check the data in the "Adaptive" section of the "About the device" section

If the problem persists, please contact Customer Service

Category E4

E4-1 ... E4-3

If you see this error code, follow these instructions:

  • Press exit
  • Restart your phone
  • Turn on the device and repeat the measurement process

If the problem persists, please contact Customer Service

E4-4


This message appears during GLUCUBE measurement.

The possible causes for this error are:

  • Loss of coverage that prevents correct synchronization with the cloud
  • Loss of internet connection that prevents correct synchronization with the cloud
Solutions for this type of error are:
  • Check if you are in a place with adequate coverage
  • Check if you have a connection good enough to be able to perform the measurement.

If you have a poor connection, you should look for a place with better connection quality and repeat the measurement

If the problem persists, please contact Customer Service.

E4-5


This message appears during the GLUCUBE measurement.

The possible causes for this error are:

  • Bluetooth disconnection from the device during measurement
  • Device disconnection during measurement
Solutions for this type of error are:

  • Check that the smartphone's Bluetooth is activated and that it does not turn off during measurement
  • Check that the device is properly charged

If the problem persists, please contact Customer Service

E4-6 ... E4-7


If this error code appears, follow these instructions:

  • Press exit
  • Restart the phone
  • Turn on the device and repeat the measurement process

If the error persists, please contact Customer Service


E5 Category


If this error code appears, follow these instructions:

  • Press exit
  • Restart the phone
  • Turn on the device and repeat the measurement process

 If the error persists, please contact Customer Service